Roll over image to zoom in
The AUTOCORR® PULSE OXIMETER looks at pulse oximetry data in a unique way instead of relying on a voltage threshold. The result is reliable readings making it a perfect solution for your most challenging environments. What’s more, this accurate, cost- effective technology is readily adaptable to clinical, home and sleep study environments.
Features and Benefits
- Reliably measure SpO2, Pulse Rate and Pulse Strength.
- Uses refined hardware and a patented algorithm to identify artifact.
- Analyzes digitized signal at a rate of 120 times per second, creating dynamic templates to validate latest signals.
- Digital, analog and nurse call output via optional cables.
- Real time or trend printouts available when used with a computer and printer.
- Remote alarm function available.
- Internal rechargeable battery.
- Genuine BCI® reusable & disposable sensors available.
Materials Included
- Autocurr® Transportable Pulse Oximeter
- 1 Adult Finger Sensor
- AC Adapter
- User Manual
Technical Specifications
Displays, Indicators & Keys
- SpO2 3-digit LED numeric display: 0.43 inch (10.9 mm) high
- Pulse Rate 3-digit LED numeric display: 0.43 inch (10.9 mm) high
- Pulse Strength: Logarithmically scaled 8-segment LED bar graph
- Low Battery LED: flashes when approximately 30 minutes of battery life remains
- Alarm Silence LED: flashes to remind the user that audio alarms are disabled
- Sensor LED: flashes to alert the operator to check the SpO2 sensor placement
- Lost Pulse Alarm: Flashing % SpO2 dashes indicate the sensor no longer detects a pulse while a finger is inserted into it (a pulse must previously be detected in order to get this alarm)
- Brightness: Adjustable brightness of SpO2, pulse rate, and bar graph displays
- Search Light: indicates the monitor is looking for a pulse
- Artifact Light: indicates the presence of excess motion or other artifact
- Power Green: color indicates monitor is connected to AC power
SpO2
- Range: 0-100%
- Accuracy: ±2% at 70-100%, ±3% at 50-69%
- Averaging: 4, 8, or 16 pulse beat average
- Alarm Range:
- >, High – 50%-100% & OFF (1% steps)
- >, Low – OFF & 50%-99% (1% steps)
Pulse Rate
- Range: 30-254 bpm
- Accuracy: ±2% at 30-254 bpm
- Alarm Range:
- High – 5-250 bpm & OFF (1 bpm steps)
- Low – OFF & 5-250 bpm (1 bpm steps)
- Averaging: 8 or 16 second average
Power
- AC Charger:
- Wall Mount 105-125V ~, 60Hz, 100V ~, 50Hz (opt)
- Table Top 230V ~, 50/60Hz (opt)
Power
- Battery: Standard Lead-Acid battery
- Battery Life: Internal rechargeable, not user replaceable. Fully charges in 6 hours. Approx. 4.5 hrs of continuous use.
Printer Output
- Data Log Mode: SpO2 and pulse rate can be printed every five seconds
- Trend Printing: Data saved every four to thirty seconds can be printed
- Optional Cables: Analog Output, Digital Output and Nurse Call
Physical
- Dimensions 3.24” H x 8.5” W x 5.5” D (21.6 cm x 8.2 cm x 14 cm)
Weight
- 30 oz. (850 grams)
Environment
- Operating Temperature: 32°F to 104°F (0°C to 40°C)
- Storage Temperature: -40°F to 167°F (-40°C to 75°C)
- Relative Humidity (non-condensing): 10-95% (storage), 15-95% (operating)
Mechanical Durability Test Results
- Vibration Testing In compliance with: IEC 68-2-6, IEC 68-2-34
- Shock Testing In compliance with: IEC 68-2-27
- Drop Testing Exceeded: UL544 testing
Standards
- EN60601-1 (IEC 60601-1), UL544, CSA Std. C22.2 No. 125-
- 1984, EN60601-1-1, EMC: EN60601-1-2
Warranty Information
- 2-year warranty