Roll over image to zoom in
Note: This item can only be shipped inside the US and to a medical or educational facility.
The Vista basic was designed to ensure simplicity in all aspects of its use: from set installation, to programming, to special mode use and alarm correction. Based on a design that has sold over 100,000 units worldwide, the Vista basic provides high reliability packaged in one of the easiest pumps to use.
Features and Benefits
- Robust design of mechanical components ensures maximum durability
- Two-year preventive maintenance schedule helps save and money from typical annual inspection requirements
- Battery life of approximately 3 years before replacement is required
- DEHP-free sets protect patients. Needle-free sets protect both the patient and the clinician
- Helpful prompts walk the user through the programming process
- Easy to use, teach and learn
- Perfect for home care settings
Materials Included
- Vista Basic Large Volume Infusion Pump
- (1) Power Cord
- (1) Mounting Clamp for IV Pole
- (1) Rechargeable Battery
- User Manual
Technical Specifications
- Dimensions
- Width: 5.5′
- Height: 9.4′
- Depth: 7.8′
- Mechanism: Linear peristaltic volumetric infusion pump
- Weight: 6.8 lbs. (3.1 kg)
- Power Requirements: 100-120 VAC, 50-60 Hz, 13 Watts
- Leakage Current: Max 30A
- Battery: 7.2V, 1.2 Ah NiCd (rechargeable)
- Battery Operating Time: 3.5 hours at 125 mL/hr
- Rate Range: 0.1 to 800.0 mL/hr in 0.1 mL increments with Vista™ tubing
- Volume to be Infused Range: 0.1 to 9999.9 mL
- Occlusion Pressure
- Variable Settings
- Low: 8 PSI
- High: 17 PSI
- Variable Settings
- KVO Rate: 1 mL/hr or selected rate, whichever is lower
- Flow Rate Accuracy: ± 5% measurement time 8hr, temp. 71.6° F
- Alarms
- Air-in-Line
- Low Battery
- Plug In Now
- Bag Empty
- Downstream Occlusion
- Door Open
- Pump Stopped
- KVO
- Unit Malfunction
- Air-in-Line Alarm
- Air bubble size 0.3mL* or 1.5mL/hr** (*Can be set from 0.01 to 0.3 mL in service program, **Can be set from 0.5 to 3.5 mL/hr in service program)
- Memory
- User option to retain last Continuous and Piggyback Mode data upon power up. Permanent data retention in Program and Ramp/Taper Modes after pump is turned off
- Tubing: B. Braun Vista™ administration sets
- Modes/Features: Continuous and Piggyback Modes, Program Mode for variable infusions, Ramp/Taper Mode for TPN infusions and Delay A Mode for postponing infusionstart and finish
- Operating Temperatures
- Operating: 64.4° F to 95° F, 10.15 PSI to 15.37 PSI
- Non-Operating: -13° F to 131° F, 10.15 PSI to 15.37 PSI
Warranty Information
- 1-Year Warranty
Documentation
User Manual
Quick Reference Guide