Infinium Omni K Patient Monitor

In stock

Compare
SKU: 1.049.998 Categories: ,

Designed for a fast paced work environment, the Infinium Omni (K)™ patient monitor offers an easy to use interface. Patient information along with vital sign settings can be quickly modified to meet the needs of a patient’s changing condition.



Features and Benefits

  • High resolution 10.5 inch LED touch screen with keypad and trim knob
  • 72 hour graphic and tabular trending
  • Wireless network compatibility
  • Four hour backup battery
  • Pre-configured with non-invasive blood pressure, 3/5 ECG with arrhythmia detection, SpO2, and temperature, respiration, OxyCrg, Drug dose calculations
  • Optional components available
    • Sidestream CO2
    • Mainstream CO2
    • Cardiac Output Invasive Blood Pressure – Includes 2 IBP cables and a password to open the IBP function
    • Embedded Thermal Printer
    • Phasin Anesthesia Multi Gas Module

Materials Included

  • Omni K Patient Monitor
  • (1) Adult Blood Pressure Cuff
  • (1) Blood Pressure Cuff Hose
  • (1) Adult Finger Sensor
  • (1) Sensor Cable
  • (1) 4 Pin Skin Surface Temperature Probe
  • (1) Power Cord
  • User Manual

Technical Specifications

  • Performance Specifications
    • Display: 10.5 inch color touch screen
    • Trace: 8 waveforms
    • Indicator:
      • Alarm indicator
      • Power indicator
      • QRS beep and alarm sound
    • Trend time: 1 – 72 hour
    • Recorder:
      • Built-in, thermal array, 3 channels
      • Record width: 48 mm
      • Recorder paper: 50 mm
      • Record speed: 25mm/s, 50mm/s
  • ECG
    • Input: 3-Lead ECG cable and standard AAMI line for connection
    • Lead Choice: I, II, III, aVR, aVF, aVL, V, V1-V6, TEST
    • Gain Choice: x0.5, x1, x2, x4
    • Frequency Characteristic: 0.05 ~ 35 HZ (+3dB)
    • ECG Waveforms: 2 channels
    • Penetration Voltage: 4000VAC 50/60Hz
    • Sweep Speed: 12.5, 25, 50 and 100 mm/sec (left to right or right to left)
    • HR Display Range: 30 ~ 300bpm
    • Accuracy: ± 1 bpm or ±1%, whichever is greater
    • Alarm Limit Range Setting: Upper limit 100 ~ 200 bpm, lower limit 30 ~ 100 bpm
  • Respiration
    • Measure Method: RA-LL impedance
    • Range: 0 ~ 120 rpm
    • Accuracy: ± 3 rpm
    • Alarm Limit Setting: Upper limit 6 ~ 120 rpm, lower limit 3 ~ 120 rpm
    • Sweep Speed: 12.5, 25, 50 and 100 mm/sec (left to right or right to left)
  • NIBP
    • Measuring Technology: automatic oscillating measurement
    • Cuff Inflating: &lt,30s (0 ~ 300 mmHg, standard adult cuff)
    • Measuring Period: AVE&lt,40s
    • Mode: Manual, Auto
    • Measuring Interval in AUTO Mode: 2 min ~ 4 hrs
    • Pulse Rate Range: 30 ~ 250 (bpm)
    • Measuring Range:
      • Adult/Pediatric Mode:
        • SYS: 40 ~ 250 (mmHg)
        • DIA: 15 ~ 200 (mmHg)
      • Neonatal Mode
        • SYS: 40 ~ 135 (mmHg)
        • DIA: 15 ~ 100 (mmHg)
    • Accuracy:
      • Maximum Mean error: ± 5 mmHg
      • Maximum Standard deviation: 8 mmHg
    • Resolution: 1 mmHg
    • Overpressure Protection:
      • Adult Mode: 300 (mmHg)
      • Neonatal Mode: 160 (mmHg)
    • Alarm Limit Setting:
      • SYS: 50 ~ 240 mmHg
      • DIA: 15 ~ 180 mmHg
  • Temperature
    • Range: 25 ~ 50 (°C)
    • Accuracy:
      • ± 0.2°C (25.0 ~ 34.9°C)
      • ± 0.1°C (35.0 ~ 39.9°C)
      • ± 0.2°C (40.0 ~ 44.9°C)
      • ± 0.3°C (45.0 ~ 50.0°C)
    • Display Resolution: 0.1°C
    • Alarm Limit Setting:
      • Upper limit 0 ~ 50°C
      • Lower limit 0 ~ 50°C
    • Channel: 2 Channels
  • IBP (Optional)
    • Measurement Range: -50 ~ 300mmHg
    • Channel: 2 channels
    • Pressure Transducer: sensitivity, 5V/V/mmHg
    • Impedance Range: 300 ~ 3000
    • Transducer Sites: ART, PA,CVP, RAP, LAP, ICP
    • Unit: mmHg/kPa selectable
    • Resolution: 1mmHg
    • Accurancy: ±1 mmHg or ±2%,whichever is greater
    • AlarmRange: -10 ~ 300mmHg
  • EtCO2 (Optional)
    • CO2 Measurement Range: 0 ~ 99mmHg
    • Accuracy:
      • ±2 mmHg (0 ~ 38 mmHg)
      • 39-99 mmHg ±5% of reading +0.08% for every 1 mmHg (above 38 mmHg)
    • Sampling Rate: 50 ml/min
    • Initialization Time: 30 seconds (typical), reaches ±5% steady-state accuracy within 3 minutes.
    • Respiration Rate: 0 ~ 150 breaths/min
    • Mode: Adult, Neonate
  • Networking: Industry standard 802.11b/g wireless network
  • Power
    • Source: External AC power or internal battery
    • AC Power: 100 ~ 240VAC, 50/60Hz, 150VA
    • Battery: Built-in & rechargeable lithium ion
    • Operating Time: 3+ hours
  • Environmental Specifications
    • Temperature:
      • Operating: 5 ~ 40 °C
      • Storage: -20 ~ 65 °C
    • Humidity Range:
      • Operating: 80 %
      • Storage: 80 %

Warranty Information

  • 2-Year Warranty