Performing spirometry tests could not be easier than with the new Micro Direct alpha spirometer. Simply enter the patient’s age, height and gender, select FVC or VC and begin the test series.
Test information is presented in an easy to read format on a clear, bright color display. The best test is automatically selected and built-in quality control checks insure accurate results. Test data along with the diagnostic interpretation and estimated lung age can be instantly printed on the integrated printer or to a PDF file via a USB connection with our report generation PC software.
The alpha features a highly accurate Fleisch pneumotach flow measuring system, battery operated portability and is shipped complete with all necessary accessories in a study carrying case.
Features and Benefits
- Rapid, low cost testing
- Accurate Fleisch pneumotach
- Color high resolution display
- Fast, quiet internal printer
- Rechargeable battery pack
- Real-time curve visualization
- On-screen help menu
- On-screen test quality prompts
- Clear, easy-to-read report
- Pre/Post bronchodilator comparison
- PDF report generation software
- Autoclaveable flow head
- Complete with durable carrying case
Materials Included
- Alpha Spirometer
- Carrying Case
- 4 Cardboard Mouthpieces
- 1 Spirosafe Filter
Technical Specifications
- Dimensions: 12” x 10” x 3”
- Parameters Measured: FVC, VC, FEV1, FEV1 ratio, PEF, F25-75, FEF25, FEF50, FEF75, FIF25, FIF50 and FIV75
- Predicted Normals: NHANES III
- Operating Temperature Range:
- ATS/ERS Limits: 63-98 F
- Design Limits: 50 – 104 F
- Voltage/Frequency: 110 – 250V ~ 50/60Hz
- Flow Detection Principle: Fleisch type pneumotachograph
- Volume Detection: Flow integration sampling @ 100Hz
- Volume Accuracy: Better than +/- 3%
- Maximum Display Volume: 10 L
- Maximum Test Duration: 90 seconds
- Flow Impedance:
- <,0.1kPa/L/S @ 14L/s
- Complies with ATS/ERS 2005
- Accuracy when Operated in Operating Temperature Range Conditions:
- Flow +/- 10%
- Max flow rate +/- 16L/s
- Min flow rate +/- 0.02L/s
- Safety Standards: EN ISO 60601-1:2005 -1 -2 -4
- Medical Safety Standard: Medical Devices Directive 93/42/EEC1993
- Designed & Manufactured Under: ISO 13485:2003, FDA 21 CFR820, CMDR & JPAL
- Communications: USB and Compact Flash
- Performance Standards: ATS/ERS 2005 & EN 13826
- Printer: Thermal
Warranty Information
- 5-year warranty
Documentation